Application Note

Mixing Efficiencies For The 100 L imPULSE Single-Use Mixer

GettyImages-2280187429-lab-with-mixing-equipment-and-scientists

Mixing performance can vary markedly with working volume and material behavior. Testing of the 100 L Thermo Scientific imPULSE Single-Use Mixer examined liquid-to-liquid and liquid-to-solid mixing from 10 L to 100 L using salt solutions, salt granules, and cell culture medium. Liquid blends achieved 95% homogeneity in less than two minutes, while sinking solids were incorporated in under eight minutes. Floating powders required longer processing times, highlighting the need to adjust agitation rates and timelines based on material characteristics.

Review the results, analytical methods, and recommended settings to support consistent mixing across your upstream and downstream bioprocess applications.