Millipore's ZipTip Cures Small Sample Prep Woes

Preparing tiny samples of biomolecules for instrumental analysis taxes the creative capabilities of most biologists. Samples going into mass spectrometers, in particular, require extra care. I've seen all sorts of gizmos used for this kind of sample prep—capillaries filled with desalting media, Pasteur pipettes plugged with alternating layers of cotton and ion exchange resins, etc. Most makeshift methods require adding buffer to recover the sample, thereby diluting the sample and sometimes adding another (concentration) step. Now there is a better way.

The new ZipTip sample preparation system from Millipore Corp. can be used to concentrate, desalt, and remove detergents from biological samples. ZipTip consists of a 10 µL pipette tip with approximately a half microliter bed of resin fixed at its end. ZipTipC18, the first of what we hope will be many products in the ZipTip series, desalts femtomoles of peptides in less than 60 seconds. Elution in 1 - 4 µl of acetonitrile/water simplifies direct transfer to a mass spectrometer.

ZipTipC18

In operation, ZipTip is placed on a standard single or multi-channel pipettor. Sample is aspirated and dispensed through the resin bed to bind, wash and elute peptides and small proteins. Supported by protocols developed at Millipore, ZipTipC18 saves time by providing a fast, ready to use device for offline sample preparation.

Table of Contents
Features and Benefits
How it Works
Reagent compatibility of ZipTipC
Specifications

Features and Benefits of ZipTipC18

  • Pipette tip device format provides a simple, easy to use Sample preparation and precise liquid handling in one device
  • C18, 15 µm beads assure rapid sample processing. According to Millipore samples are purified in sixty seconds
  • Robust: well-characterized interaction Elution in solvent/matrix for direct delivery to a MALDI target
  • Suitable for use with single, multi-channel and automated liquid handling systems
  • Flexible Increased throughput saves time
  • Small 0.6 microliter chromatographic bed volume is suitable for minute sample volumes and masses Minimal sample loss

How it Works

The following protocol is used to prepare samples using ZipTipC18 for sample binding, washing, and elution of peptides or small proteins (< 40 Kd) for MALDI MS, Electrospray MS, or other analytical techniques.

Materials: ZipTipC18 contains C18 spherical silica (15 µm , 200Å pore size) in a 0.6 µL bed volume; ZipTipC18 pipette (single or multi-channel); wetting buffer: 50% acetonitrile in water; equilibration buffer: 0.1% triflouroacetic acid (TFA) in water; elution buffer: 50% acetonitrile in water. For direct MALDI, prepare matrix in 50% acetonitrile/water.

NOTE: Since the adsorptive bed provides back pressure, do not use ZipTip for accurate volumetric dispensing. To achieve optimal performance, adjust the P-10 pipette to 10 µL and press ZipTip on securely. Depress plunger to dead stop to allow maximum suction and sample dispensing.

Multiple pipette ZipTips save time by processing multiple samples

Protocol: Use the ZipTip with samples containing picogram to micrograms of sample. The absorption efficiency of ZipTip is sample- and concentration- dependent. Maximum binding to the ZipTip is achieved in the presence of TFA or other ion-pairing agents. To minimize sample dilution and enhance analyte binding, use 10% TFA to achieve a final concentration of 0.1% TFA. In the case of excess detergent, dilute sample with 0.1% TFA to achieve acceptable binding conditions, for example, SDS (< 0.1%), Triton (< 1%), and Tween (< 0.5%).

To equilibrate the ZipTipC18 for Sample Binding

  1. Prewet the tip by depressing plunger to a dead stop using the maximum volume setting of 10 µL. Aspirate Wetting buffer into tip. Dispense to waste. Repeat.
  2. Equilibrate the tip for binding by washing it twice with the Equilibration buffer. (See "Materials" section for details.)

To bind, wash, and elute peptides after ZipTip equilibration:

  1. Bind peptides and proteins to ZipTipC18 by fully depressing the pipette plunger to a dead stop. Aspirate and dispense sample 5 to 10 cycles depending on sample concentration. Dilute solutions require increased contact time.
  2. Wash tip and dispense to waste using two cycles of 0.1% TFA, water, or 5% methanol in 0.1% TFA/water mixture. You may require additional washing for electrospray applications or samples containing high salt or detergent.
  3. Elute by dispensing 2 to 4 µL of Elution buffer to clean vial using a standard pipette tip. Carefully, aspirate and dispense eluant through ZipTip at least three times without pushing air into the sample. For Electrospray MS, elute sample into clean vial and apply directly for analysis. For MALDI MS, elute with or without matrix in 50% acetonitrile/water.

    CAUTION: Acetonitrile is volatile and evaporation can occur rapidly. If this occurs, add more eluant to recover sample.

    Pipette 1 to 2 µL of desalted-concentrated sample directly onto target by depressing plunger until appropriate volume is dispensed. Save or discard the remaining sample with tip.

    NOTE: For samples containing 1 µg of protein or 0.5 µg of peptide, you can achieve maximum recovery by increasing elution volume to 10 µL or by performing multiple elutions (2 × 5 µL). For information on fractionating a peptide, see the "Technical Assistance" section.

    Protein recovery yields using ZipTip are excellent

    Back to the Top

    Table 1. Reagent compatibility of ZipTipC18

    Compatibility code:
    YES - Long exposures at room temperature have no significant effect.
    ? - Questionable - Short exposures at room temperature cause little or no damage.
    No - Not recommended - Short exposure may cause permanent damage.

    Acetic Acid (Glacial)YES
    AcetoneNO
    Mercaptoethanol (0.1M)YES
    Acetonitrile (100%)YES
    Aliphatic Esters?
    Ammonium Hydroxide (5%)YES
    Benzene (100%)NO
    Butyl AcetateNO
    Chloroform (1%)?
    Dichlorobenzene (100%)NO
    Dichloromethane (1%)?
    Diethanolamine (5%)YES
    Dimethyl Acetamide (100%)NO
    Dimethyl Formamide (1%)YES
    Dimethyl Formamide (100%)NO
    Ethanol (100%)YES
    Formic Acid (5%)YES
    Guanidine HC1 (6 M)YES
    Hydrochloric AcidYES
    Hydrogen PeroxideYES
    Isopropyl Alcohol (100%)YES
    Mercaptoethanol (1.0M)?
    Methyl Alcohol (100%)YES
    Methyl Ethyl Ketone (100%)NO
    Methyl Isobutyl Ketone (100%)YES
    n-Butanol (100%)YES
    Nitric Acid (0.1N)YES
    Nitric Acid (1.5N)YES
    Phenol (0.5%)NO
    Phosphoric Acid (1M)YES
    Sodium Azide (1%)YES
    Sodium Hydroxide (0.1N)YES
    Sodium Hydroxide (0.5N)YES
    Sodium Hypochlorite (100 ppm)YES
    Sodium Hypochlorite (200 ppm)YES
    Sulfuric Acid (1%)YES
    Toluene (1%)NO
    Triton X-100YES
    TweenYES
    Urea (6M)YES

    Triton X-100 is a registered trademark of Union Carbide Chemical and Plastics, Inc. Tween is a registered trademark of ICI Americas, Inc.

    Back to the Top

    Specifications

    Materials of Construction: Pipette tip: Polypropylene Resin: C18 spherical silica (15 µm, 200Å pore size)
    Tip capacity: 10 µL Adsorptive bed: 0.6 µL Length: 31 mm Max. capacity: 2 µg when used with saturating amounts of analyte
    Max. Temperature: 70°C Min. Temperature: 4°C
    Working pH Range: 2 to 13.

    By Angelo DePalma

    For more information: Donna Murphy, Technical Services, Millipore Corp., 80 Ashby Road, Bedford, MA 01730-2271. Tel: 781-533-8244.